Pour water into the exhaust bottle.
Add water to the exhaust bottle
as
described below.
Drain po
Pol
Reference
line
level
(1)
UnIoad the exhaust bottle from the body.
Pull
the bottle outwards until
the
top
handle
can
be
grabbed
securely.
Lift
the bottle out of the autoclave using
this
handle.
(2)
Pour water into the bottle through the water
filling
port.
Fill
water to the reference line level.
LOW
level
1
Drain
valve
(3)
Check to make sure that the water level is at the
LOW
level (the
lowest water level).
If
too much water has been poured in, then place the bottle in a
level sink with the side of the water filling and drain ports facing
downwards. Any excessive water is drained automatically until
the water level is lowered to the
LOW
level.
(4)
Check to see that the drain valve, located at the bottom of the
exhaust bottle housing
area,
is closed.
(5)
Load
the
bottle into the area.
Be sure to push the bottle to the end, or else
an
error
(
E
,
E
)
will
occur.