When the pressure
in
the working chamber falls below
0.1
kgf/cm2, the exhaust valve is
fully
opened.
When the temperature in the working chamber (or the temperature resistered
by
the floating
sensor,
if
used) falls to
97°C
(or 7g°C)*, the
EWWT.
display stops flickering and lights up. The
display of opening of the exhaust valve (exhaust
%)
disappears from the digital display, and
operation proceeds to the next cycle.
*
Specifications
for
particular
area.
Forced Exhaust Cycle
---
Mode
3
The
IEXWT,
display goes out and starts flickering, and the automatic exhaust valve is fully
opened.
When the temperature in the working chamber falls to
97"C, the
IEXWT.
display stops flickering
and lights up, and operation proceeds to the next cycle.
II
Heating Cycle
---
Mode
4
The
ST-BY
display stops flickering and lights up, and the
WIEATG
display goes out and starts
flickering. The digital display indicates the temperature in the working chamber.
Temperature rises until a set liquefaction temperature is reached.
After the set temperature is reached,
a
set liquefaction time is shown on the digital display upon
the activation of the liquefaction
timer.
The digital timer displays the remaining time during the heating cycle. Refer to "Checking Set
-
Values during Operation" on page 19 for the method of checking the set time during operation.
When the time preset for liquefaction has passed, the HEATING LED stops flickering and lights
up, and operation proceeds to the next cycle.
Warming Cycle
---
Modes
1
and
4
The
WARM
display goes out and starts flickering.
When the temperature in the working chamber falls to a set incubation temperature, the electronic
alarm gives
a
beeping sound.
When
20
hr.
(fixed) have passed after the temperature drops to the set incubation temperature, the
WARM
display stops flickering and.lights up, and operation proceeds to the next cycle.
-A
NOTE
After the warming time (20
hr.)
has passed, the working chamber is not heated; temperature in
the chamber falls to room temperature, and the remaining agar medium, if any, in the chamber
will be coagulated.
a
When you take a sterilized substance out of the working chamber during
the
warming
cycle, press
the
START6TOP
switch to stop operation. Refer
to
"9.
Unloading"
on page 21.