If steam is abruptly exhausted after sterilization of liquid, the liquid may gush out. To prevenl
this, set
the opening of the exhaust valve (Exhaust
%)
to a small value to gradually exhaust.
I
Or, set that to
0%
(spontaneous cooling).
When used with
a
sterilization can, it takes several hours for the temperature in waste disposal
bag to reach the set temperature after the temperature in the
working chamber (displayed
temperature)
reaches the set value (time lag).
If there is approx.
300
-
500
milliliters
of
water in the waste disposal bag, steam is generated
in
the bag and drives the
air
out. This will significantly reduce the time lag at the time of
temperature rise. Refer to the table below
and
take this time lag into account when setting the
fertilization time.
Model:
HV-110
Case:
A
large number of 015
x
100
test tubes placed in a waste disposal bag
77
HV-110
Reference Values for Time Lag
in
Bag
0
milliliters
of
water
For liquefaction of coagulated agar medium, set an appropriate liquefaction temperature and
time,
referring to the
table below.
HV-50
Reference Values (per
Flask)
Mode, temperature, time and opening
of
exhaust valve remain in the memory by means of the
backup battery even after the power switch'is turned off. When the backup battery has run
down, the
CLOCK
display flickers, and the initial set values for Mode 1, described in the
previous section, come back on display. In this case, call the authorized sales outlet from which
you purchased the product.
Quantity of Liquid
2
liters
1
liter
500
cc
Liquefaction Temperature
100
"C
Liquefaction Time
60
minutes
45
minutes
25
minutes