9.
Unloading
A
WARNING
Check to see that the gauge for pressure in the working chamber reads
0
kgf/cm2.
-
A
CAUTION
Do not put your face or hands close to the working chamber when lifting the lid after
operations are complete; steam will gush out
of
the opening of the chamber.
After operation is over, the lid, chamber,
packing and panel are hot. To protect yourself
from
bums, do not touch with bare hands.
It
takes a lot of time for the liquid to cool. Be sure to check that the temperature has dropped
sufficiently before unloading
the
liquid from the working chamber, or you may get burned.
Put on heat insulation gloves before taking a substance out of the
working chamber.
Do not
put your hands into the chamber until the steam has been vented.
O
Slide the open/close lever to the
UNLOCK
side (to the right).
O
Lift
the lid.
O
Take the sterilized substance out of the
working chamber.
10.
After Completion of Operations
O
Turn off the power switch after the completion of each routine operation.
A
IMPORTANT
To prevent clogging of the piping, change water within the working chamber once daily,
r
refemng to
"Draining
Working
Chamber"
on page
27.