11.
To
Interrupt Operation
A
NOTE
If
operation is interrupted during the sterilization of liquid, the liquid may overflow into the
worlung chamber.
I
O
Press the
STARTISTOP
switch.
The ongoing process will be interrupted, and the equipment returns to the
standby state (state before operation).
When you take the sterilized substance out of the working chamber, follow
the instructions described in
"9.
Unloading".
(When chamber temperature
drops below
97'C (or 7g°C)* in the
SOLID
mode, pressure drops to
0
kgfl
cm?.
In this
state, the
opedclose lever can be unlocked.)
(If
a
floating sensor is used, the sensor regulates the open/close lever
by
sensing the temperature.)
*
In
LIQ
mode available in specifications for particular
area.
12.
If
Power Supply
Is
Cut
Off
during Operation
.If
the power supply is cut off due to power failure or the like, operation is interrupted. When
power supply is restored, the equipment will be back in the standby state (state before operation).
In this case, repeat operations from the beginning.
-
,---
A
NOTE
1
0
If power supply is cut off due to power failure or the like, the opedclose lever is locked for
safety. To open or close the lid, follow the instructions described
in
"9.
Unloading"
on page
21
after the power supply has been restored.
13. Operation
of
Cycles
Illumination
circled
Turn-on Timer On Cycle
---
Common to a11 the modes
(with the turn-on timer set)
The digital display section illuminate in a circular
manner until the time set on the turn-on timer (preset
time
of
operation start-up) is reached.
When the time set on the turn-on timer is reached, the
TIMER
digital goes off, and operation
proceeds to the next cycle.
Air
Evacuation Cycle
---
Mode
1,2
and
3
The
ST-BY
display stops flickering and lights up; then the
BIEATG
goes off and start
flickering. The temperature in the working chamber is displayed in the digital temperature
display section.