Do not pile specimens on top of one another. When the chamber is overly packed, steam fails
to penetrate
to
all points, resulting in incomplete sterilization.
0
In sterilizing liquids such as chemical and medium, pay attention to the quantity
of
the liquid
in relation
to
its container. For
an
Erlenmeyer
flask,
the amount of chemical should
be
approx.
314
of the capacity of the container; for
a
test tube, the appropriate quantity of chemical is
approx. half of the capacity of the container. Too much chemical may result in overflow from
the container during the temperature rising or cooling process.
When using a container, loosen the cap or use
a
breathable cap. An unbreathable cap
may
cause the container to burst.
In
the case of liquefaction of agar medium, its quantity should be 2 liters or less per container.
Two liters
or
more of agar medium ,may not be completely liquefied.
4.
Selecting
Mode
(Process)
The following modes are programmed in the microcomputer. Select a mode suitable for your
purpose.
O
Press the
MODE
switch.
Each time you press the switch, the current mode repeatedly changes from
\.
/
Mode
1
to Mode 2,
3,4,
1,
...
in this order.
Mode
1
2
Application
Sterilization of
agar
medium (warmed for the prevention of coagulation after sterilization).
Sterilization of liquid, such as water, medium, reagent, liquid medicine, that withstand
high
3
temperature, high pressure steam.
Sterilization of tools of glass, ceramic, metal or rubber
that
withstand high temperature,
high
4
pressure steam and abrupt de pressurization during an exhaust process.
Liquefaction of medium.